Recall Z-0804-2016 ยท BSZ Gas-Machine, Anesthesia

Maquet Cardiovascular Us Sales, Llc

Flow-i Anesthesia System Product Usage: The indication for the Flow-i Anesthesia System is administering inhalation anesthesia while controlling the entire ventilation of patients with no ability to breath, as well as in supporting patients with a limited ability to breath. The system is intended for use on neonatal to adult patient populations. The system is intended for use in hospital environm

DateFeb 22, 2016 (initiated Dec 10, 2015)
StatusTerminated
ReasonElectrical issues in the control, monitoring or panel sub-systems, caused by the printed circuit boards (PCBs) that control the sub-systems. The problems manifested as multiple re-starts of the sub-systems and, in a few instances, disabling of the system has occurred. Replacement PCBs are now available to correct the issues. Only the Model C20 is affected in the US.
Root causeNonconforming Material/Component
ActionMaquet sent "Urgent Medical Device Field Action Field Correction For Flow-i Anesthesia Systems" letters/Field Correction Response forms dated 12/10/2015 via UPS to affected customers. The notification informed customers of the issue, asked them to distribute the notification to appropriate staff; verify the serial number(s) and informed them they could continue to use the system if they were not experiencing electrical issues. Customers were informed that a Maquet service representative would be contacting them to arrange a time to replace their affected units. Customers were asked to com
Quantity7 units within US
DistributionUS Nationwide in the states of CA, NY, and OR
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133958FLOW-I ANESTHESIA SYSTEM C20, C30 & C40
Maquet Critical Care AB, Mar 20, 2014