ClearanceBrief
Recall Z-0897-2016 ยท GCQ Endoscope, Flexible

Integrated Medical Systems Inc

Olympus, Model No. PCF-160AL, flexible scope, large diameter, video colonoscope

DateJan 20, 2016
StatusTerminated
ReasonIf excessive pressure is applied to the distal end of the endoscope, the internal coil pipe assembly at the distal end of the endoscope can become detached and angulation may not operate as intended.
Root causeNonconforming Material/Component
ActionThe firm notified their consignees of the issue by Fed Ex on 01/20/2016. The firm plans to coordinate with the medical facilities for inspection of the endoscopes.
Quantity5 instruments
DistributionDistributed US (nationwide) and in Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.