Recall Z-1171-2016 ยท LDP Colorimetry, Acetaminophen

Sekisui Diagnostics P.E.I. Inc.

Sekure Chemistry Sekisui Diagnostics Acetaminophen L3K Assay, In-vitro Diagnostic Catalog Number: 506-10: R1: 1 x 10ml, R2: 2 x 10ml Catalog Number: 506-30: R1: 3 x 10ml, R2: 6 x 10ml In-Vitro quantitative measurement of Acetaminophen in serum and plasma used m the diagnosis and treatment of acetaminophen overdose toxicity.

DateJan 22, 2016
StatusTerminated
ReasonCrystal formation in the Color Reagent (R2) when left on board automated analyzers causing samples to under recover
Root causeDevice Design
ActionSekisui sent an Important Field Correction - Customer Notification letter dated January 22, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. 1. Discard all reagent in house with a current expiration date of 2016-09. 2. Update the expiration on all remaining kits on hand with the new expiration dates provided 3. When adding additional reagent to the analyzer, a new reagent wedge should be used. 4. If crystals are observed within reagent stored on-board, reagent should be discarded and replaced with fresh reagent.
Quantity14,407 kits.
DistributionWorldwide Distribution - US Distribution to the states of : AZ, CA, FL, IA, ID, IL, MA, , MN, MT and ND., and to the countries of Canada, France and Hong Kong.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081938ACETAMINOPHEN L3K ASSAY, AND ACETAMINOPHEN L3K ASSAY, MODELS 506-10, 506-30
Genzyme Diagnostics P.E.I., Inc., May 1, 2009