Recall Z-1020-2016 ยท KPS System, Tomography, Computed, Emission

Philips Medical Systems (Cleveland) Inc

BrightView are gamma cameras; Designed for single or dual detector nuclear imaging.

DateDec 17, 2015
StatusTerminated
ReasonThe firm discovered four issues with the BrightView Family systems. Issue 1: During a cardiac non-Auto Body Contouring (ABC) scan, using Cardiac High Resolution (CHR) collimators, the collimator may come in contact with the bottom edge of the patient pallet. If this occurs, collision sensors would pause the scan. Issue 2: During a scan, the gantry motion may halt prematurely. This cause
Root causeSoftware design
ActionOn 12/22/2015, the firm sent Customer Information Letters to their consignees. The letter identified the product, problem and actions to be taken by the customers.
Quantity677 units
DistributionWorldwide Distribution-US (nationwide) including the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, VT, WA, WI, WV & WY, and in DC and PR and in the countries of Algeria, Ar
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062298APOLLO GAMMA CAMERA SYSTEM
Adac Laboratories, Aug 23, 2006