Recall Z-0752-2016 ยท OWB Interventional Fluoroscopic X-Ray System

Toshiba American Medical Systems Inc

INFX-8000V Bi-Plane X-Ray Interventional System X-ray systems

DateFeb 9, 2016 (initiated Aug 14, 2015)
StatusTerminated
ReasonWhen a fontal DA (Digital Angiography) acquisition was done, scattered x-ray came into the dose meter on the lateral side. As a result, the dose meter sent a minus value to the software. Consequently, the software defined the data as an "abnormal value" and it stopped displaying dose data and the dose data was lost. The following message was displayed, "Dose meter abnormal, Dose info disabled".
Root causeRadiation Control for Health and Safety Act
ActionToshiba's planned action to bring into compliance: 1.Affected customers will be notified using the Customer Notification Letter you provided, 2. revised system software will be installed on the systems to prevent this occurrence from happening, 3. the corrections will be made free of charge, and 4. the corrections will be completed by April 4, 2016. For further questions, please call 1- 800 521-1968.
Quantity4
DistributionUS Nationwide Distribution to OH and NY
FDA recordOpen on fda.gov