ClearanceBrief
Recall Z-1243-2016 ยท CHP Radioimmunoassay, Estradiol

Siemens Healthcare Diagnostics, Inc

IMMULITE 2000 EstradiolTest Code E2 Catalog Number: L2KE22, L2KE22 (D), L2KE26, L2KE26 (D) SMN: 10381178, 10702833, 10381177, 10702834 Product Usage: For in vitro diagnostic use the IMMULITE2000 systems Analyzers- for the quantitative measurement of estradiol in serum, as an aid in the diagnosis and treatment of various sexual disorders

DateJan 13, 2016
StatusTerminated
ReasonThe drug fulvestrant (Faslodex) may cause falsely elevated estradiol results.
Root causeDevice Design
ActionSiemens issued an Urgent Medical Device Correction (UMDC) to affected Siemens Healthcare Diagnostics customers in the United States on January 13, 2016 to be delivered to customers on January 14, 2016 via Federal Express. An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on January 13, 2016. The letters identified affected product, problem and the actions to be taken. Customers may continue to use the estradiol assays to report patient results however, the assays should not be used for patients being treated with fulvestrant. Cust
Quantity9334 units
DistributionWorldwide Distribution US Nationwide , AR, AU, AU, BD, BO, BR, CA, CH, CL, CN, CO, DO, EC, EG, HK, ID, IE, IL, IN, JP, KR, LK, LY, MX, MY, NI, NP, NZ, PE, PH, PY, SA, SG, SV, TH, TW, UY, VE, VN, and ZA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K932926IMMULITE ESTRADIOL
Diagnostic Products Corp., Jan 10, 1994