Recall Z-0797-2016 ยท HXC Wrench

Orthofix, Inc

Firebird Spinal Fixation System Torque Limiting Handle (PN 52-1512) Intended for posterior, non-cervical pedicle, and non-pedicle fixation (T1-S2/llium).

DateFeb 11, 2016 (initiated Jan 8, 2016)
StatusTerminated
ReasonThere is a possibility that the device may deliver a torque value less than the required setting after initial sterilization. In worst case, the unexpected drop in torque value may lead to the spinal construct loosening and requiring corrective surgical intervention.
Root causeComponent design/selection
ActionAffected consignees were notified via an Urgent Medical Device Recall Notification letter, dated 1/8/16. The letter identified the affected device and stated the reason for the recall. The recalling firm requests customers to remove the affected devices from their inventory and cease further distribution or use. Customers whom the devices were further distributed to should be notified that the affected devices cannot be used and must be returned to Orthofix per the instructions provided. The attached Acknowledgement Form should be completed and returned. Customers can contact their local
Quantity36 units
DistributionWorldwide Distribution -- US, Spain, Italy, and Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151488Firebird Spinal Fixation System/Phoenix MIS Spinal Fixation System
Orthofix, Inc., Jul 7, 2015