Recall Z-0975-2016 ยท PIF Gastrointestinal Tubes With Enteral Specific Connectors
Xeridiem Mediem Medical Devices Inc
EndoVive 3s Low Profile Balloon Kits Part Number M00548430 (XMD P/N 70-0050-A14) The EndoVive 3s Low Profile Balloon is intended to provide gastric access through a gastrointestinal stoma tract for enteral feeding, medication administration and gastric decompression. The EndoVive 3s Low Profile Balloon is indicated for use in adult and pediatric patients who require long-term feeding, are unable t
| Date | Dec 23, 2015 |
| Status | Terminated |
| Reason | A feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric contents to leak out of the device in some instances. |
| Root cause | Device Design |
| Action | Xeridiem sent a customer notification letter dated December 23, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customer was provided the following instructions: 1. Segregate all devices from the above table that in Boston Scientific Corporation's inventory and restrict them from being distributed to end users. 2. Determine the list of end users who have received devices from the above list, notify them of the quality issue, and have these devices returned to Boston Scientific Corporation. Provide Xeridiem Medical |
| Quantity | 36 |
| Distribution | Nationwide distribution to MA only. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K142297 | Endo Vive 3S Low Profile Balloon Kit, Endo Vive 3S Bolus Extension Sets, Endo Vive 3S Continuous Extension Sets, Endo Vi Xeridiem (Formerly Mri), Jan 8, 2015 |