Recall Z-0862-2016 ยท LLZ System, Image Processing, Radiological

Siemens Medical Solutions USA, Inc

Syngo Plaza Picture archiving and communication system. Software only. PACS intended to display, process, read, report, communicate, distribute, store and archive digital medical images. Stores and archives within DICOM structured reports. Integrates hospital/radiology information systems.

DateFeb 23, 2016 (initiated Dec 22, 2015)
StatusTerminated
ReasonSiemens is releasing an updated software version to address several software issues including RGB images will show "?" since calculation of HU is not possible; save as option enabled; changes in access for loading studies; breast region is now properly fitted to segment boundary when clicking fit breast to screen.
Root causeSoftware design
ActionSiemens sent a Voluntary Correction and Removal Action letter dated December 30, 2015. The letter identified the product, the problem, and the action to be taken by the customer. Customers were notified that a software update VB10A_HF01/06 is being released simulataneously to eliminate the issue. Customers with questions were instructed to call 1-888-826-9702.
Quantity68
DistributionNationwide Distribution including AL CA CT FL IN KS LA MA MD MI MN MO MS NE NY OH OR PA TX WA WI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K093612SYNGO.PLAZA
Siemens AG Medical Solutions, Mar 4, 2010
K132532SYNGO.PLAZA VB10A
Siemens AG Healthcare, Dec 9, 2013