Recall Z-0683-2016 ยท IYE Accelerator, Linear, Medical

Iba Dosimetry Gmbh

IBA Dosimetry GmbH Lift Table Type 4320, Catalog HA03-010#2 and HA03-000#2

DateJan 21, 2016 (initiated Dec 10, 2015)
StatusTerminated
ReasonWhen the electric lift table is connected to the mains power, in a very rare occasion, the table may start to move up in an uncontrolled manner.
Root causeComponent design/selection
ActionThe field safety notice including a confirmation of receipt was sent to the affected hospitals; by email on 12/10/2015 then followed by 12/17/2015 sent by Fed Ex. The firm requested the mains power be disconnected after setting up the table and BP. Replacement part and instructions will follow.
DistributionNationwide and International: Algeria, Angola, Arab Republic of Egypt, Austria, Belarus, Belgium, Bulgaria, Bosnia and Herzegovina, Croatia, Czech Republic, Denmark, Ethiopia, France, Germany, Ghana, Greece, Hungary, Iraq, Ireland, Islamic Republic Iran, Italy, Kazakhstan, Kenya, Kingdom of Saudi
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K945321WELLHOFER WP700 V. 3.2
Medical Physics Instrumentation, Inc., Jul 12, 1995