Recall Z-0738-2016 ยท CHH Enzymatic Esterase--Oxidase, Cholesterol

Horiba Instruments Inc

The ABX PENTRA Cholesterol CP ref. A11A01634 is an in vitro diagnostic assay for quantitative determination of cholesterol in human serum and plasma based on an enzymatic photometric test (Trinders reaction). The assay is composed of a 99 ml mono-reagent cassette. Reagent is a chemical solution with additives.

DateFeb 2, 2016 (initiated Jan 11, 2016)
StatusTerminated
ReasonN-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.
Root causeUnder Investigation by firm
ActionCustomer notification letters were sent on 01/11/16. Customers were provided the following instructions/actions: When performing tests that use the Tinder reaction, the blood draw should be done before treatment is given to the patient. Reagent inserts for the concerned HORIBA Medical Products have been updated and now include the following additional warning: Patients treated with N-Acetylcysteine (NAC) for Paracetamol overdose may generate a false low result. The letter request customers share the information with laboratory staff and retain the notification as part of their laboratory do
Quantity216
DistributionU.S. distribution to the following; CA, NY, AL, OK, MN, MS, MD, TX, KY, WV, MT, IL, FL, VA, WA, MI, KS, NE, PA, OH, ID, OR, AR, AZ, NM, NC, SC, GA, TN No foreign distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060854ABX PENTRA 400: LIPOPROTEINS
Horiba Abx, Aug 24, 2006