Recall Z-0745-2016 ยท IYE Accelerator, Linear, Medical

Philips Medical Systems

DS/US Proton Feature with Pinnacle3 Software Version 10.0.0 and 14.0.0 Radiation Therapy Planning System Model #459800200621

DateFeb 2, 2016 (initiated Jan 8, 2016)
StatusTerminated
ReasonIn DS/US proton planning, there is a correction factor used internally to the dose engine that is calculating improperly in some cases. This can result in an incorrect dose calculation.
Root causeDevice Design
ActionPhilips sent an "Urgent Medical Device Correction" letter dated 16-DEC-2015 to affected customers. The letter described the Affected Products, Problem Description, Hazard Involved, How to Identify Affected Products, Action to be Taken by Customer / User, Actions Planned by Philips and Further Information and Support (contact Philips Customer Care Solutions Center at 1-800-722-9377 ).
Quantity3
DistributionUS: Nationwide Distribution in the states of FL, OH, and MO.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K130992PINNACLE3 RADIATION THERAPY PLANNING SYSTEM
Philips Medical Systems (Cleveland), Inc., Jun 14, 2013