Recall Z-0675-2016 ยท FPA Set, Administration, Intravascular

Rymed Technologies, LLC

InVision-Plus and InVision-Plus Junior Needleless IV Connector, Disposable, Single Use Products are intended for single patient use in intravenous and blood administration sets without the need for needles, thus eliminating the potential for needle-stick injuries during use.

DateJan 21, 2016 (initiated Jan 5, 2016)
StatusTerminated
ReasonPotential for the female luer component of the InVision-Plus IV connector to crack during use.
Root causeProcess design
ActionThe firm Ry Med Technology LLC, sent an "URGENT: MEDICAL DEVICE RECALL" letter dated January 4, 2016 and response form to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to discontinue use and return the product to: RyMed Technologies LLC, 6000 W. William Cannon Drive, Building B, Suite 300 Austin, TX 78749; the user facilities to change the IV connectors in 24 hour when lipids are being administered; if further distributed, contact other accounts, advise of recall and return outstanding stock; complete and return the RECALL NOTIF
Quantity496,400 units
DistributionUS Nationwide Distribution.