Recall Z-1244-2016 ยท IYE Accelerator, Linear, Medical

MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK

MHI-TM2000 Linear Accelerator System (Software Version 3.0.0 and after) Intended for radiation therapy of lesions, tumors. conditions anywhere in the body where radiation therapy is indicated.

DateMar 24, 2016 (initiated Dec 15, 2015)
StatusTerminated
ReasonDue to Operator Console software anomaly, a change in treatment completion status of the last patient of the day may be altered from "Completed (or Discontinued)" to "Untreated" under specific conditions.
Root causeSoftware design
ActionDue to Operator Console software anomaly, a change in treatment completion status of the last patient of the day may be altered from "Completed (or Discontinued)" to "Untreated" if all the following conditions are met: (1) At least one patient has been treated on the day; (2) A major error occurs when VeroTM/MHI-TM2000 is operated in a mode, other than Clinical Mode, which means Daily Check Mode or Maintenance Mode. (3) The major error is NOT associated with a function that uses therapeutic/ MV radiation parameters, such as an error in Isocenter Check function, etc. Note 1: If the major erro
Quantity25 devices (4 domestically in U.S.A. and 21 internationally)
DistributionDistributed in the states of Florida, New York, Ohio & Texas, and the countries of France, Germany, Japan, Italy, Korea, & Belgium.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072047MHI-TM2000 LINEAR ACCELERATOR SYSTEM
Mitsubishi Heavy Industries, Ltd., Aug 29, 2007
K122450MHI-TM2000 LINEAR ACCELERATOR SYSTEM VERO IMAVIS
Mitsubishi Heavy Industries, Ltd., Oct 3, 2012
K152867MHI-TM2000 Linear Accelerator System, VERO, Imavis
Mitsubishi Heavy Industries, Ltd., Oct 21, 2015