Recall Z-0689-2016 ยท NIO Stent, Iliac

Cook Medical Incorporated

Zilver 518RX Vascular Stent with Rapid Exchange Delivery System

DateJan 22, 2016 (initiated Dec 22, 2015)
StatusTerminated
ReasonManufacturing of the joint between the inner catheter and stylet of the delivery system does not consistently meet specifications, and the joint may separate during deployment. Failure of this joint during a procedure may result in a partial stent deployment. The clinical effect of a partial deployment may be a need for surgical intervention, or may result in vessel trauma or an embolic event
Root causeDevice Design
ActionCook Medical sent an "URGENT: MEDICAL DEVICE RECALL" letter dated December 22, 2015, to all affected customers. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Should you have any questions or concerns, please do not hesitate to contact Cook Medical Customer Relations at 1-800-457-4500 or 1-812-339-2235. On 2/5/2016 a second recall notification was sent to affected consignees, expanding the recall to All Lots of the affected product.
Quantity753
DistributionNationwide Worldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.