Recall Z-0691-2016 ยท CBK Ventilator, Continuous, Facility Use
Draeger Medical, Inc.
Emergency Transport Ventilator Oxylog 2000 Plus Ventilator Catalog Number: 5705080. Volume controlled emergency and transport ventilator with pressure support.
| Date | Feb 3, 2016 (initiated Dec 22, 2015) |
| Status | Terminated |
| Reason | Drager became aware of situations where the error message "Poti unplugged" was generated causing acoustic and visual alarms, the breathing system releases pressure and the ventilation function stops operating. |
| Root cause | Device Design |
| Action | The firm, Draeger Medical, Inc., sent an "Urgent Medical Device Recall" letter dated 12/21/2015 and a copy of the Instructions for Use (IFU) Supplement to consignees in a method that confirms delivery. The letter described the product, problem and actions to be taken. The consignees were instructed to include a copy of the supplement with their Oxylog IFU; inform all affected users in their facility; and contact Draeger Customer Support at 1-800-543-5047 (press 4 at the prompt) to obtain additional copies of the IFU Supplement Free of Charge. If there are any questions regarding this letter, |
| Quantity | 0 |
| Distribution | Nationwide Distribution -- including states of AR, AZ, CA, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. |