Recall Z-0744-2016 ยท DXT Injector And Syringe, Angiographic

Acist Medical Systems

ACIST Medical Systems, BT2000 Automated Manifold Kit (SKU 014613) and CPT2000 (SKU 014623) Automated Manifold Kits, Sterile R, Rx Only, Made in Mexico. The manifold body is identical in the BT2000 and CPT2000 kits. The BT2000 and CPT2000 kits facilitate connection to, and delivery of, either contrast or saline into the patient via a coronary catheter (not provided by ACIST). To be used for th

DateFeb 2, 2016 (initiated Dec 21, 2015)
StatusTerminated
ReasonReports related to ingress of air into the manifold.
Root causeNonconforming Material/Component
ActionAcist sent an Urgent Medical Device Recall letter dated December 21, 2015, via e-mail to all affected consignees. The letter was addressed to Distribution Partner or Cardiac Cath Lab / Material Manager in the USA. In Europe, the letter was addresses to Purchasing Department. The letter described the problem and the product involved in the recall. Advised consignees to return affected units and to complete and return the Medical Device Recall Form. Distributors were further advised to send the recall letter to their customers. For questions or concerns, contact in US, ACIST Customer Servic
Quantity48,614 cartons or 486,140 units
DistributionWorldwide Distribution - US including all states except DE, MT, RI, UT, & VT and Internationally to Belgium, France, Germany, Iceland, Ireland, Luxembourg, Netherlands, Reunion, United Kingdom, Australia, Austria, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Estonia, Gre
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052744C2000 AUTOMATED MANIFOLD KIT
Acist Medical Systems, Inc., Dec 16, 2005
K984231ANGIOGRAPHIC CONTRAST MANAGEMENT SYSTEM, MODEL PART A AND PART B
Invasatec, May 31, 2000