Recall Z-0818-2016 ยท NBH Accessories, Arthroscopic
Smith & Nephew, Inc., Endoscopy Div.
High Flow Irrigation Extender; Part Number 72201021. Arthroscopic accessories are indicated for use in access of the joint capsule.
| Date | Jul 30, 2013 |
| Status | Terminated |
| Reason | During arthroscopic surgery, when the scope is moved inside the cannula, the body of the irrigation extender separates, potentially obstructing visualization and causing leakage of the irrigation fluid. The device is removed and a back-up device is used to complete the surgery. |
| Root cause | Under Investigation by firm |
| Action | Urgent - Product Recall 1st Notification - Urgent R-2013-13 letters, dated July 30, 2013 were sent overnight via Federal Express to customers. The letters identified the affected product, as well as the potential risk involved. Actions to be taken were provided for both hospital and sales representatives. Hospital representatives were asked to inspect their inventory, locate all devices from the list provided, and quarantine them immediately. The Inventory Return Certification Form should be completed and returned, regardless of whether any affected product is on hand or not. Contact the |
| Quantity | 203 units |
| Distribution | Worldwide Distribution -- US, including the states of AL, AR, CA, FL, GA, ID, IL, KS, MA, MI, MO, NC, TN, TX, and AK; and, country of Japan. |
| FDA record | Open on fda.gov |