ClearanceBrief
Recall Z-0722-2016 ยท IZI System, X-Ray, Angiographic

Philips Electronics North America Corporation

INTEGRIS H5000; Model Number: 72246 The Philips INTEGRIS 3000 I 5000 system is intended for interventional and diagnostic vascular and neurovascular procedures. The Philips INTEGRIS Allura 9 Biplane system is intended for diagnostic cardiovascular, vascular and interventional procedures. The Philips INTEGRIS Allura 12 & 15 system is intended for peripheral, abdominal, cerebral diagnostic and inte

DateJan 30, 2016 (initiated Nov 3, 2015)
StatusTerminated
ReasonThe use of Allura Xper, Allura Clarity, Allura CV, Allura Centron and UNIQ systems regarding live images and still images can lead to still images being interpreted as live Images.
Root causeDevice Design
ActionThe firm, Philips, issued "URGENT-Field Safety Notices" dated 10/26/2015 and 11/19/2015 to its customers. The notices described the product, problem and actions to be taken. The customers were instructed to ensure the audible signal (X-ray buzzer) is always enabled, to activate only ONE fluoroscopy pedal at time, and to configure the footswitch to the default Philips configuration. Customers were also instructed to review the information with all members of your staff who need to be aware; retain a copy with the equipment IFU and complete and return the Confirmation Respond form to Mail Stop 4
Quantity13297 in total
DistributionWorldwide Distribution: US (Nationwide) including Puerto Rico and countries of: Albania, Algeria, US Virgin Islands, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Demo
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K973482PHILIPS INTEGRIS V5000 AND BV5000
Philips Medical Systems North America, Inc., Nov 10, 1997