Recall Z-0567-2016 ยท JKA Tubes, Vials, Systems, Serum Separators, Blood Collection

Becton Dickinson & Company

BD Vacutainer PLUS K2EDTA Tubes (Plastic). 13 x 75 mm x 2.0 mL (Pink) BD Hemogard Additive: K2EDTA (spray dried), 3.6mg Sterile IVD BD, Made in USA The Vacutainer Brand PLUS (Plastic) Tube with EDTA Vacutainer Brand Serum Tube are evacuated blood collection tubes which provide a means of collecting, transporting, separating and processing blood in a plastic tube.

DateDec 31, 2015 (initiated Dec 3, 2015)
StatusTerminated
ReasonSome of the Blood Collection Tubes in these lots were manufactured with insufficient EDTA additive. An insufficient EDTA additive may lead to erroneous platelet results, specimen recollection, or a delay of treatment.
Root causeProcess change control
ActionThe firm, BD, issued an "URGENT PRODUCT RECALL" notifications/Recall Response Forms to their customers dated 12/3/2015 via email and/or UPS 2nd Day Mail. The notification identified the reason for the recall; how to identify affected product; and actions to be taken. The customers were instructed customers to immediately review their inventory; complete the enclosed Recall Response Form and fax (1-866-873-0312 it to BD or email it to bd4964@stericycle.com; and return all affected products with the completed Recall Response Form following instructions on the enclosed packing instruction, even
Quantity29,815,600 units in total
DistributionWorldwide Distribution: US (Nationwide) and the countries of: European Union, Peoples Republic of China, Japan and Republic of Korea.