Recall Z-0467-2016 ยท LNH System, Nuclear Magnetic Resonance Imaging

Siemens Medical Solutions USA, Inc

MAGNETOM systems, magnetic resonance diagnostic devices (MRDD).

DateDec 22, 2015 (initiated Dec 7, 2015)
StatusTerminated
ReasonPossible installation error of the quench lines of superconducting magnets, used to vent helium gas in the rare event of a quench. Any restriction, obstruction, or improper installation of the quench line may cause the helium gas to be blown directly into the magnet room or other areas. This could lead to displacement of oxygen, cause cold burns, and/or pose a risk to users, patient or others.
Root causeOther
ActionSiemens sent an Customer Advisory letter dated December 7, 2015, certified mail to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter notifies customers of the potential injuries, which may result of improperly installed quench lines. The letter indicates it is the system's owner who is responsible for installing the quench line per the requirements. If Siemens installed the quench lines, the consignee is directed to contact Siemens, who will inspect the quench lines during next maintenance visit. Siemens will
Quantityin US 3752 units
DistributionNationwide including : DC, PR, Virgin Islands and Bermuda
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002179MAGNETOM ALLEGRA SYSTEM
Siemens Medical Solutions USA, Inc., Oct 11, 2000
K032428MAGNETOM AVANTO MR SYSTEM
Siemens Medical Solutions USA, Inc., Oct 16, 2003
K130885MAGNETOM AVANTO-FIT, MAGNETOM SKYRA-FIT
Siemens Medical Solutions USA, Inc., May 17, 2013
K132951MAGNETOM AERA
Siemens Medical Solutions USA, Inc., Nov 15, 2013
K133226BIOGRAPH MMR
Siemens Medical Solutions USA, Inc., Nov 12, 2013