Recall Z-0630-2016 ยท CGA Glucose Oxidase, Glucose

Arkray Factory USA, Inc.

arkray SPOTCHEM II Glucose Reagent test strips, Item No. 77101, MEDTEST DX Catalog # 77424. For the in vitro quantitative determination of glucose in human plasma from whole blood samples.

DateJan 15, 2016 (initiated Dec 18, 2015)
StatusTerminated
ReasonOne lot each of SPOTCHEM II Basic PANEL-1 Reagent Test Strip and SPOTCHEM II Glucose Reagent Test Strip are being recalled due to a lowered reaction rate in the reagent pad for glucose. The affected lots will display a result of approximately 300 to 400 mg/dL for any glucose value that exceeds the upper limit of 416 mg/dL.
Root causeComponent design/selection
ActionThe firm sent an Urgent Medical Device Recall letter to customers beginning 12/18/2015. The letter identified affected product, stated the reason for recall and asked for unused product to be returned. For instructions on returning unused product contact Lesty Jaranilla 1-877-538-8872 x3243.
Quantity88 boxes (25 foil packaged test strips per box)
DistributionDistributed in the states of FL, IL, KY, MI, NC, NY, OH, and TN.