Recall Z-1043-2016 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Aesculap, Inc.

Columbus Revision Knee System, EnduRo Knee System Product Usage: The Columbus Revision Knee System and EnduRo Knee System are indicated for use in reconstruction of the diseased knee joint.

DateNov 30, 2015
StatusTerminated
ReasonAesculap Inc. US has initiated a recall on Tibial and Femur extension sterile pressfit stem implant packaging that are used for Columbus Revision/Enduro Knee Implants because it is labeled as Cementless. US product is indicated in the US for use with Bone Cement only.
Root causePackaging change control
ActionAesculap sent an Urgent Medical Device Correction letter to Distributors and Sales Rep. The letter identified the affected product, problem and actions to be taken. Customers were notified to add the over label to product packaging. Telephone conferences with Distributors and Sales Rep reiterated need for representative to inform Surgical Handler of the label when providing product for review. Customers were instructed to complete the attached Product Correction Acknowledgement Form and return to Aesculap Quality Assurance department by faxing the form to 610-791-6882 or e-mail to val.straw
Quantity1232
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K083772COLUMBUS REVISION KNEE SYSTEM
Aesculap Implant System, Inc., Jun 4, 2009