Recall Z-0550-2016 ยท LLZ System, Image Processing, Radiological

Siemens Medical Solutions USA, Inc

Syngo Imaging XS is a Picture Archiving and Communication System (PACS)

DateNov 17, 2015
StatusTerminated
ReasonFor the Syngo Workflow SLR System with Software Ver: VB10C: Printouts may be printed in incorrect anatomical size when using syngo Imaging XS filming application in conjunction with a printer not released for anatomical print usage. This may happen due to an invalid combination of printer and the syngo Imaging XS printer configuration file.
Root causeSoftware in the Use Environment
Action11/17/2015, CSAN sent to customers with a list of suitable printers for the anatomical printing.
Quantity42
DistributionAL ; CA ; CO ; CT ; FL ; GA ; IA ; IL ; IN ; KS ; MD ; MI ; MO ; NC ; NE ; NJ ; NV ; NY ; OH ; PA ; TN ; TX ; WI ;
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082430SYNGO IMAGING XS
Siemens AG Medical Solutions, Nov 14, 2008