ClearanceBrief
Recall Z-0471-2016 ยท HWQ Passer

Biomet, Inc.

Black Mamba Suture Passer

DateDec 2, 2015
StatusTerminated
ReasonThere may be a separation and an insufficient weld between the spring and washer. The weld between the spring and washer may fracture, causing the Mamba Suture Passer instruments jaw to not actuate properly. 6 complaints have been received in which 3 have been reported with a delay up to 10 minutes.
Root causeNonconforming Material/Component
ActionOn 12/2/2015, URGENT MEDICAL DEVICE RECALL NOTICE notifications were sent to the affected distributors and hospital risk managers via courier with instructions to locate and return the affected products. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.
Quantity50
DistributionFL, NJ, WI, UT, TX, KY, KS, OH, GA, LA, AL, NY, NC, AR, MO, VA, CA, MN, IN, HI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.