Recall Z-0453-2016 ยท HAB Saw, Powered, And Accessories
Zimmer, Inc.
Zimmer UNIVERSAL Power System Loaner and Modular Electric/Battery Single Trigger Handpiece. Rx only Made in Switzerland
| Date | Dec 21, 2015 (initiated Nov 10, 2015) |
| Status | Terminated |
| Reason | The firm identified a malfunction which made it possible for a handpiece to start by itself when the power source is connected. |
| Root cause | Device Design |
| Action | Zimmer sent an Urgent Medical Device Correction letter dated November 9, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. 1. Review this notification. 2. You will receive copies of notification sent directly to hospital risk managers and surgeons in your territory. As necessary, review and facilitate understanding of this notification by those entities. 3. Further identify hospital risk managers and/or surgeons in your territory who should also receive notification of this update, and supply the information to t |
| Quantity | 89 units |
| Distribution | US Distribution to the states of : AR, CA, CO, CT, FL, GA, HI, IL, KS, KY, LA, ME, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, PA, RI, SC, TX, UT, VA, WA, WI, WV and WY. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.