Recall Z-1044-2016 ยท LCQ Fluorescent Immunoassay Gentamicin

ConvaTec, Inc

Nebulizer Masks Opti-Mist and Opti-Mist Clear Nebulizer masks are intended to spray liquids in aerosol form into gasses that are delivered directly to a patient.

DateMar 3, 2016 (initiated Nov 11, 2015)
StatusTerminated
ReasonThe Nebulizer mask malfunctions in that there is reduced or no aerosol output. The malfunction occurs from the reduction in the diameter of a component part inside the nebulizer.
Root causeNonconforming Material/Component
ActionThe firm, ConvaTec, sent each consignee affected by the recall an overnight express mail "Urgent Voluntary Recall Notice" and Response Form dated 12/7/2015. The notice described the product, problem and actions to be taken. The recall notice instructed the consignee to immediately stop distributing and quarantine all affected lots; perform a count of all recalled product currently in inventory; complete and return the enclosed Recall Response Form by email at convatec3657@stericyle.com or by fax at 1-888-603-8986, and return all affected product, (ONLY return the lots affected with this recal
Quantity3,605,700 pieces
DistributionWorldwide Distribution: US (nationwide) including states of: AL, AZ, AR, CA, CO, CT, FL, GA, IL, IN, IA, KS, ME, MA, MN, MS, MO, NV, NH, NJ, NY, NC, OH, OR, PA, SC, TX VA, WA, WI, and countries of: Australia, Belgium, Burma, Canada, Denmark, Egypt, Estonia, Faroe Islands, Finland, France, Germany, I
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K801867DISPOSABLE MEDICATION NEBULIZER
Hospitax, Aug 20, 1980