Recall Z-0637-2016 ยท JJJ Counter (Beta, Gamma) For Clinical Use

Perkinelmer

Perkin Elmer WIZARD2 10-detector, 1000 samples. Product Code: 2470-0200, 3470-0200. The Wizard gamma counter is intended to detect and count gamma radiation emitted by clinical samples

DateJan 12, 2016 (initiated Dec 8, 2015)
StatusTerminated
ReasonThe content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect. The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement protocols have been programmed for both ID #023 and #024, the protocol associated with barcode ID #024 is executed. If the error is undetected, the Gamma Counter may produce erroneous results.
Root causeNonconforming Material/Component
ActionPerkin Elmer sent an Urgent Medical Device Correction letter dated December 7, 2015 to affected customers. The letter identified the affected products, problems and actions to be taken. Customers were instructed to destroy all affected products in their possession. Customers were asked to complete the enclosed Correction Response Form and return it by scan and e-mail to QMresponse.SG@perkinelmer.com . Corrected barcode labels ID #023 will be provided upon request and return of the completed Correction Response Form to PerkinElmer. For further information contact your local PerkinElmer repre
Quantity22 units
DistributionWorldwide Distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, KUWAIT, MYANMAR, NETHERLANDS, NEW ZEALAND, NORWAY, PO
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.