Recall Z-0450-2016 ยท HRS Plate, Fixation, Bone

Biomet, Inc.

A.L.P.S. 4.0mm Cancellous Locking Screws intended for use in stabilization and fixation of fractures, revision procedures, fusions, reconstructions (osteotomy) and non-unions of the bones.

DateDec 18, 2015 (initiated Nov 20, 2015)
StatusTerminated
ReasonThese screws have a single lead (instead of a triple lead) thread. The screw may sit proud in the plate, If the screw does not lock into the plate; the screw may back out. If the screw sits proud, soft tissue irritation may occur. If the screw does not lock into the plate and backs out, a revision surgery may be necessary.
Root causeProcess change control
ActionOn 11/20/2015, URGENT MEDICAL DEVICE RECALL NOTICE notifications were sent to the affected distributors, hospital risk managers, and implanting surgeons via courier. Distributors were instructed to locate and return the affected product, hospitals to locate and discontinue use, and implanting surgeons to monitor patients. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Questions related to this notice should be directed to (574) 372-1570, Monday throug
Quantity182, 965
DistributionAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR CHILE, COSTA RICA, NETHERLANDS, CANADA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072083SMALL FRAGMENT LOCKING PLATING SYSTEM
DePuy Orthopaedics, Inc., Oct 25, 2007
K072832ANTEROLATERAL AND MEDIAL LOCKING PLATING SYSTEM
DePuy Orthopaedics, Inc., Dec 14, 2007
K082300ANATOMIC LOCKING PLATING SYSTEM
DePuy Orthopaedics, Inc., Feb 18, 2009
K111663TRAUMA INTERNAL FIXATION SYSTEMS
DePuy Orthopaedics, Inc., Mar 12, 2012