Recall Z-0647-2016 ยท MUJ System, Planning, Radiation Therapy Treatment

Brainlab AG

RT Elements are applications for radiation treatment planning for use in stereotactic, conformal, computer planned, Linac based radiation treatment of cranial, head and neck, and extracranial lesions. The simulated plan is intended for treatment evaluation for example in tumor board meetings or operating rooms

DateNov 5, 2015
StatusTerminated
ReasonLarge objects with fine resolution are potentially displayed cropped when imported into Brainlab Brain Metastases 1.0.0.
Root causeSoftware design
ActionBrainlab sent a FIELD SAFETY NOTICE / PRODUCT NOTIFICATION letter dated November 5, 2015, to all direct accounts informing the of the recall and actions to be taken. User Corrective Action: Users of Brain Metastases Version 1.0.0 or Adaptive Hybrid Surgery Analysis 1.0.0 shall adhere to the following: After import always carefully review and verify all objects for correctness and validity. For reviewing an object, select the image set which was used to outline or to modify the objects in the Data menu. Also compare the contours of an object in Brain Metastases or Adaptive Hybrid Surgery Analy
Quantity20 systems
DistributionWorldwide Distribution - US (nationwide) and Internationally to Australia, Belgium, Brazil, Canada, China, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Mexico, Portugal, South Korea, Spain, Switzerland, Turkey and the United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142108RT ELEMENTS, DOSE REVIEW, BRAIN METASTASES, ADAPTIVE HYBRID SURGERY ANALYSIS
Brainlab AG, Dec 16, 2014