Recall Z-0434-2016 ยท GKZ Counter, Differential Cell
Beckman Coulter Inc.
UniCel DxH 600 Coulter Cellular Analysis System, Software Version 1.1.1.0, Part No. B23858 The DxH Slidemaker Stainer is a fully automated slide preparation and staining device that aspirates a whole-blood sample, smears a blood film on a clean microscope slide, and delivers a variety of fixatives, stains, buffers, and rinse solutions to that blood smear.
| Date | Dec 16, 2015 (initiated Dec 1, 2015) |
| Status | Terminated |
| Reason | Beckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creation of multiple orders with the same Specimen identification (ID) but different Patient identification when manually editing pending orders at the System Manager. |
| Root cause | Software design |
| Action | Beckman Coulter sent an Urgent Medical Device Recall letter dated December 1, 2015, to all affected customers to inform them that Beckman Coulter is recalling the the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because of the possibility of sample misidentificaton. Customers were informed that during internal testing, Beckman Coulter determined that the software for the DxH systems noted above allows the creation of multiple orders with the same Specimen identification (ID) but different Patient identification when manually editing pending orders a |
| Quantity | 666 units total (327 units in US) |
| Distribution | Worldwide Distribution - US (nationwide) and Internationally to Australia, Austria, Belarus, Belgium, Bosnia and Herzegovina, Brunei Darussalam, Canada, China, Colombia, Croatia, Czech Republic, Ecuador, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jo |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K081930 | UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM Beckman Coulter, Inc., Dec 19, 2008 |
| K120771 | UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM Beckman Coulter, Inc., Mar 22, 2013 |
| K140911 | UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM/ UNICEL DXH SLIDEMAKER STAINER COULTER CELLULAR ANALYSIS SYSTEM Beckman Coulter, Inc., Sep 5, 2014 |