ClearanceBrief
Recall Z-0760-2016 ยท DWJ System, Thermal Regulating

C.R. Bard, Inc.

Arctic Sun 5000 and Arctic Sun 2000; Intended for monitoring and controlling patient temperatures.

DateFeb 5, 2016 (initiated Nov 10, 2015)
StatusTerminated
ReasonImproper targeted temperature therapy was delivered to patients because the Temperature Simulator Key was connected to the patient temperature cable instead of the patient temperature probe.
Root causeUse error
ActionBard Medical Division sent a Safety Alert to each of the customers via FedEx with proof of delivery notice. All consignees have been instructed to carry the notification out to the user-level.
Quantity1,539
DistributionNationwide Distribution-DC and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI. IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071341MODIFICATION TO ARCTIC SUN TEMPERATURE MANAGEMENT SYSTEM, MODEL 2000
Medivance, Inc., Aug 3, 2007
K101092ARCTIC SUN TEMPERATURE MANAGEMENT SYSTEM, MODEL 5000
Medivance, Inc., Jul 16, 2010