Recall Z-0597-2016 ยท LLZ System, Image Processing, Radiological

Siemens Medical Solutions USA, Inc

Software version VD10E for Syngo X-Workplace; Picture archiving and communication system.

DateJan 6, 2016 (initiated Nov 18, 2015)
StatusTerminated
ReasonPotential post-processing software issue when using tabcard "4D" on X-Workplace with software version VD10E. A too small measurement in the MPRs of a volume when images are acquired with a CT scanner with a tilted gantry. In a 2x2 layout, the 4D tabcard will show an incorrect, too small length measurement. This could result in selecting a device of the wrong size, which then needs to be exchanged.
Root causeSoftware design
ActionSiemens sent an Important Customer Safety Notice dated November 17, 2015. The Notice identified the product, the problem, and the action to be taken by the customer. The potential issue will be remedied with a software update via Update Instructions AX067/15/S. Siemens service organization will contact customers to arrange a date for the installation of the software update. Customers with questions were advised to call 800-888-7436.
Quantity73 units
DistributionNationwide Distribution including: AL, AZ, CA, CO, FL, IA, ID, IL, IN, MA, MD, MI, MN, MO, MT, NC, NY, OH, OK, PA, SC, TN, TX, VA, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K143319Syngo X-workplace SW VD10
Siemens Medi Cal Solutions, Inc., Feb 24, 2015