Recall Z-0480-2016 ยท HAW Neurological Stereotaxic Instrument

Synaptive Medical, Inc.

Synaptive Medical BrightMatter Navigator System (Guide System). Part number SYN-0026. The Guide system is comprised of a cart labelled SYN-0024 and a cart labelled SYN-0025, Operator Cart. A planning and intraoperative guidance system to enable open and percutaneous computer assisted surgery. The system is indicated for medical conditions requiring neurosurgical cranial procedures. It can be us

DateDec 23, 2015 (initiated Dec 4, 2015)
StatusTerminated
ReasonOut of tolerance for radio frequency emissions. At the 150-1000MHz frequency, the testing indicated the BrightMatter Navigation system was up to 20dB uV/meter higher than the applicable IEC 60601-1-2:2007 (Ed3.0) standard specification.
Root causeComponent design/selection
ActionConsignees were sent via courier on 12/4/2015 a Synaptive "Urgent Medical Device Recall" letter dated 30 November 2015. The letter described the problem and the device involved in the recall. Informed consignees that a technical support team member from Synaptive Medical will conduct the field correction. They also requested consignees to complete and return the Recall Response Form. For questions contact Thas Yuwaraj, PhD, .PEng Monday through Friday, 9 AM to 5 PM, Eastern Time at 1-844-462-7246.
Quantity8
DistributionUS: FL, WI, IL, LA, WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142024BRIGHTMATTER NAVIGATION SYSTEM
Synaptive Medical, Inc., Apr 2, 2015