Recall Z-0481-2016 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Artis systems with large display from Siemens Medical Solutions USA, Inc. Interventional Fluoroscopic X-ray for angiographic procedures.

DateDec 24, 2015 (initiated Oct 30, 2015)
StatusTerminated
ReasonDue to improper soldering in a limited number of large display bypass module components of a specific production lot, a loss of video signal could potentially occur. This has not been observed in the field and only sporadic cases have been observed in factory screening.
Root causeSoftware design
ActionSiemens sent an Important Customer Safety Notice dated October 29, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The issue will be remedied with a hardware update AX025/15/S. The affected bypass module hardware will be exchanged with unaffected production lots to eliminate the problem. For further questions call: (610) 219-6300.
Quantity17
DistributionUS Distribution to the states of : IL, WI, MA, FL, AL, TN, ND, OH, TX, CA, NJ and MI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K073290ARTIS ZEE AND ZEEGO
Siemens Medical Solutions USA, Inc., Feb 11, 2008
K123529ARTIS Q AND Q.ZEN- MODULAR ANGIOGRAPHIC SYSTEM
Siemens Medical Solutions USA, Inc., Feb 26, 2013