ClearanceBrief
Recall Z-0459-2016 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Siemens Healthcare Diagnostics Inc.

Reagent Management System (RMS) Dimension EXL with LM STM Product Usage: The Reagent Management System (RMS) is a module for use with the Dimension clinical chemistry system. The RMS doubles the available storage of refrigerated Flex reagent cartridges on the Dimension system from 44 to 88. The RMS loads, hydrates, and automatically removes Flex reagent cartridges, allowing the Dimension system

DateDec 22, 2015 (initiated Nov 10, 2015)
StatusTerminated
ReasonSiemens has identified RMS refrigeration compressors that are missing the safety cover on the electrical termination block. The exposed electrical hazard on the electrical termination block is located directly behind the RMS Waste Container.
Root causeNonconforming Material/Component
ActionSiemens sent an Urgent Medical Device Correction (UMDC # DI-16-01.A.US) and Urgent Field Safety Notice (UFSN # DI-16-01.A.OUS) letters dated November 2015 to all affected customers. The letters identified the affected product, problem and actions to be taken. For questions contact your local Siemens technical support representative.
Quantity4
DistributionUS Nationwide Distribution: FL, NY, SC, TX, LA, AZ, WI, IA, IL, OH, AL, AK, NH, OK, MD, CA, NM, MS, KY, NJ, MN, GA, CO, MA, SD, TN, MO, PA, MT, SC, NE, KS, VA, AL, NV, WY, OR, PR, MO, OK, GU, HI, AR, ME, VT, CT, NC and the countries of Canada and Mexico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K130276DIMENSION EXL WITH LM SYSTEM
Siemens Healthcare Diagnostics, Inc., Mar 22, 2013