Recall Z-0447-2016 ยท IYE Accelerator, Linear, Medical

Elekta, Inc.

MOSAIQ Oncology Information System. Used to manage workflows for treatment planning and delivery.

DateNov 24, 2015
StatusTerminated
ReasonFirst Databank (FDB) is a third party product incorporated into MOSAIQ. FDB routinely modifies drugs in their database to reflect market and formulary changes. In FDB Framework v2.0, those drugs are not being included in drug-drug, drug-allergy and duplicate drug therapy interaction checking.
Root causeSoftware design
ActionElekta sent an Urgent Important Field Safety Notice 371-01-MSQ-004 to all affected customers on 11/24/2015. The notice informs users to the issue, the work around and the solution. The customers are instructed to complete and return the Important Field Safety Notice Acknowledgement form to their local Elekta Office or Representative, as soon as possible but no longer than 30 days from receipt.
Quantity211
DistributionNationwide Distribution -- AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WY, & Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141572MOSAIQ ONCOLOGY INFORMATION SYSTEM
Impac Medical Systems, Inc., Sep 5, 2014