Recall Z-0464-2016 ยท OAD Catheter, Percutaneous, Cardiac Ablation, For Treatment Of Atrial Flutter
Boston Scientific Corp
IntellaTip MiFi XP Asymmetric (N4) Curve 7/110/2.5/8-8 Material Number: M004PM4500N40; Catalog number: PM4500N4; Cardiac: The IntellaTip MiFi XP Catheter is indicated for interruption of accessory atrioventricular conduction pathways associated with tachycardia, for treatment of AV nodal reentrant tachycardia, for treatment of atrial flutter tachycardia, and for creation of complete AV block in
| Date | Nov 20, 2015 |
| Status | Terminated |
| Reason | During routine monitoring of complaints, BSC noted that the IntellaTip MiFi XP N4 complaint rate for distal tip bends/kinks was above the anticipated rate in BSCs product risk documentation. |
| Root cause | Device Design |
| Action | Boston Scientific sent an Urgent Medical Device Removal letters to all affected customers dated November 20, 2015 via Fed Ex or other equivalent service. The letter identified the affected products, problem and actions to be taken. Customers were instructed to contact their local Sales Representative to answer any questions. |
| Quantity | 1556 total - all models |
| Distribution | Worldwide Distribution - US Nationwide in the state of: PA, FL, IN, MI, TN, AL, IL, KY, MI, TX, OH, MO, MA, MN, ID, AL, TN, VA, CA and internationally to Austria, Australia, Belgium, Switzerland, Germany, Great Britain, Greece, Hong Kong, Israel, Iran, Italy, Japan, Kuwait, Latvia, Malaysia, Netherl |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.