Recall Z-0482-2016 ยท GDW Staple, Implantable

Cardica, Inc.

MicroCutter Xchange 30 Blue Cartridge, FG-025320; The MicroCutter XCHANGE 30 Blue Cartridge and MicroCutter XCHANGE 30 Blue Curved Tip Cartridge contain 316L stainless steel staples. The cartridges are used with the MicroCutter XCHANGE 30 Stapler, which is a single patient device. When the MicroCutter XCHANGE 30 Stapler is loaded with a MicroCutter XCHANGE 30 Cartridge, two double rows of 316L sta

DateDec 24, 2015 (initiated Nov 6, 2015)
StatusTerminated
ReasonCardica, Inc. has received reports that use of the Blue Cartridges with the MicroCutter XCHANGE 30 Stapler was associated with incomplete firing of staples, potentially resulting in an incomplete transection or anastomosis of tissue during the surgical procedure.
Root causeProcess design
ActionAn Urgent: Medical Device Removal letter, dated November 17, 2015, was sent via Fed Ex. The letter identified the affected device, described the reason for the recall, and provided the instructions to be followed. Customers were asked to check their inventory, and remove and quarantine any affected product. The notice should be forwarded to relevant personnel, and the affected product should be returned along with the completed Reply Form. A credit will be issued for any returned product. A sales representative can be contacted with questions.
Quantity568 boxes or 6816 cartridges
DistributionWorldwide Distribution -- USA, distributed to the states of Alabama, California, Colorado, Florida, Georgia, Iowa, Massachusetts, Maryland, Minnesota, New York, Ohio, Pennsylvania, Rhode Island, Texas, Washington, and Wisconsin ; and, to the countries of Austria, Belgium, France, Germany, Ireland, I
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K132581MICROCUTTER XCHANGE 30, MICROCUTTER XCHANGE 30, BLUE STAPLE CARTRIDGE, BLUE STAPLE CARTRIDGE
Cardica, Inc., Jan 7, 2014
K140118MICROCUTTER XCHANGE 30 STAPLER, MICROCUTTER XCHANGE 30 BLUE CARTRIDGE
Cardica, Inc., Feb 25, 2014
K140170MICROCUTTER XCHANGE 30 STAPLER, MICROCUTTER XCHANGE 30 WHITE CARTRIDGE
Cardica, Inc., Feb 19, 2014