Recall Z-0412-2016 ยท FLL Continuous Measurement Thermometer

Bestmed, LLC

Digital Temple Thermometer labeled under: Bestmed, Good Neighbor (Amerisource Bergen), Kroger, Medline, Meijer, Premier Value, Safeway, Life Brand, Target, Top Care, Best Choice, Western Family. Device is packaged in plastic blister with cardboard insert, 2 AAA batteries installed, a Quick Start Guide, and an Instruction Manual booklet. Product Usage: Device is a handheld thermometer that employs

DateDec 16, 2015 (initiated Nov 12, 2015)
StatusTerminated
ReasonAn incorrect calibration of the thermometer that causes the device to display inaccurate and frequently lower temperatures than the actual body temperature of the user.
Root causeProcess control
ActionConsignees were notified verbally about 11/12/15. Written notification is planed and will include instructions on the identification of the Affected Device, potential health risk, method for return, and manner of restitution for the cost of the Affected Device.
Quantity305,548
DistributionUS Nationwide Distribution and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K103617DIGITAL FOREHEAD THERMOMETER
K-Jump Health Co., Ltd., Apr 4, 2011