Recall Z-0527-2016 ยท JXM Enzyme Immunoassay, Benzodiazepine

Ameditech Inc

ProScreen 6 Panel Dip Card, Item No. PSCuPSD-6BUP300, PSD-6MTDBO300, PSD-6BUPO, PSD-6MBO, PSD-6MB,

DateNov 16, 2015
StatusTerminated
ReasonAmeditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
Root causeNonconforming Material/Component
ActionA Medical Device Labeling Correction letter was sent on 11/16/15 to distributors and end users to inform them that Ameditechs Drugs of Abuse Tests have shown reduced reactivity for up to two (2) of the nineteen (19) BZO compounds for which specificity information is provided in the product insert. The letter informs the customers that the particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide. Customers with questions are instructed to contact Ameditech at (844) 616-4453 or email at amt.fieldaction@alere.com.
Quantity2,675 kits
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052115FIRST CHECK MULTI DRUG CUP 12 FOR MARIJUANA (THC), COCAINE (COC), AMPHETAMINE (AMP), METHAMPHETAMINE (MET), ECSTASY
First Check Diagnostics, LLC, Dec 29, 2005
K113046AMEDITECH IMMUTEST MILTI-DRUG SCREEN
Ameditech, Inc., Nov 10, 2011
K113501AMEDITECH CHECKCUP MULTI DRUG TEST (FLAT/ROUND CUP)
Ameditech, Inc., Dec 5, 2011