Recall Z-0589-2016 ยท DWC Controller, Temperature, Cardiopulmonary Bypass

Maquet Cardiovascular Us Sales, Llc

Maquet HCU 30 Coolant Type R134A Filling weight 250 g 120 V 50-60 Hz 1920 W Product Usage: The Heater Cooler Unit (HCU) 30 is intended to circulate water through heat exchange circuits to warm or cool a patient during short duration cardiopulmonary bypass procedures lasting 6 hours or less.

DateJan 5, 2016 (initiated Nov 13, 2015)
StatusTerminated
ReasonFailure of the HEATER-COOLER UNIT (HCU 30) Type 2 power supply board. A drop of more than 10% voltage can cause overheating and subsequent failing of the R56 thermistor of the power supply board. Board failure typically occurs when the compressor is switched on (prior to patient use); however, it can occur during patient use. This can shut down the compressor or the entire HCU 30 Type 2 unit.
Root causeDevice Design
ActionMAQUET sent "Product Recall Urgent Medical Device Field Correction" notices and "Medical Device Field Correction Response Forms" dated 11/13/2015 via UPS to affected customers. The notification identified the issue and how to identify the affected unit. The notification instructed customers to always keep a replacement heater-cooler unit on standby in the event of a unit failure; complete the attached Field Correction Response form (fax to 1.973.629.1518 or email to HCU30@maquet.com; and that they would be contacted by a representative of the Maquet Service Team to schedule an onsite service
Quantity100 Units (distributed in United States)
DistributionUnited States Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031544HEATER COOLER UNIT, MODEL HCU 30
Jostra AB, Aug 7, 2003