Recall Z-0430-2016 ยท JQP Calculator/Data Processing Module, For Clinical Use

bioMerieux, Inc.

MYLA CLI V3.X TO V4.1 DL380 SERVER Product Usage: MYLA is a computer application (Middleware) based on Web 2.0 technology which: Interfaces between the instruments connected to the application and the LIS(s) (Laboratory Information System(s)).

DateNov 3, 2015
StatusTerminated
ReasonMYLA software connected to a Laboratory Information System (LIS) and a VITEK 2 system has the potential to link a test result to an incorrect patient record with the same specimen ID.
Root causeSoftware design
ActionbioMerieux sent an Urgent Product Correction Notice letter dated November10, 2015 via FedEx to affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact your local bioMrieux Clinical Customer Service organization at (800) 682-2666.
Quantity626 systems in total
DistributionWorldwide Distribution - US Nationwide in the states of: AK, AL, AR, AZ, CA, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and countries of: Foreign accounts: Argentina, Australia, Austria, Belgium,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.