Recall Z-0427-2016 ยท JQP Calculator/Data Processing Module, For Clinical Use
bioMerieux, Inc.
MYLA CLINIC PATCH 3.3.0 CD Product Usage: MYLA is a computer application (Middleware) based on Web 2.0 technology which: Interfaces between the instruments connected to the application and the LIS(s) (Laboratory Information System(s)).
| Date | Nov 3, 2015 |
| Status | Terminated |
| Reason | MYLA software connected to a Laboratory Information System (LIS) and a VITEK 2 system has the potential to link a test result to an incorrect patient record with the same specimen ID. |
| Root cause | Software design |
| Action | bioMerieux sent an Urgent Product Correction Notice letter dated November10, 2015 via FedEx to affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact your local bioMrieux Clinical Customer Service organization at (800) 682-2666. |
| Quantity | 626 systems in total |
| Distribution | Worldwide Distribution - US Nationwide in the states of: AK, AL, AR, AZ, CA, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and countries of: Foreign accounts: Argentina, Australia, Austria, Belgium, |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.