Recall Z-0420-2016 ยท CCW Laryngoscope, Rigid

Verathon, Inc.

GlideScope Video Laryngoscope (GVL), Part Number GVL 3 - 0574-0007; GVL 4 - 0574-0001; GVL 5- 0574-0030. The system consists of a video laryngoscope and portable video monitor. The GlideScope Video Laryngoscopes are intended for use by qualified medical professionals to obtain a clear, unobstructed view of the vocal cords for medical procedures. GVL is a video laryngoscope that incorporates a hig

DateDec 15, 2015 (initiated Nov 30, 2015)
StatusTerminated
ReasonCertain models and serial numbers of GlideScope GVL and AVL laryngoscope blades may separate, due to inadequate bonding of the silicone adhesive affixing the halves of the blade. The potential for blade separation may not be readily visible during routine inspection before or after intubation.
Root causeProcess control
ActionThe Urgent Medical Device Recall letter, dated November 30, 2015, had an error on the dates of manufacture for the AVL devices. Once Verathon discovered the incorrect information, the notification letter was revised and it was sent on December 7, 2015 to any customers that received the previous letter between 11/30/15 and 12/2/15. **************************************************************************************** Verathon sent an Urgent Medical Device Recall letter dated November 30, 2015, to all affected consignees. Verathon will, at no cost to you, test your GlideScope GVL and/or
Quantitytotal 612 units (394 units in the US and 218 units outside the US)
DistributionWorldwide Distribution - US (nationwide including Puerto Rico) and Internationally to Bahrain, Belgium, Bermuda, Brunei Darussalam, Bulgaria, Canada, China , Czech Republic, Denmark, France, Georgia, Germany, Greece, India , Indonesia, Iran , Israel, Italy , Japan, Kuwait, Lebanon, Malays
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.