ClearanceBrief
Recall Z-0708-2016 ยท OWB Interventional Fluoroscopic X-Ray System

Philips Electronics North America Corporation

UNIQ FD; Model Numbers: 722028 The Philips UNIQ is intended for use on human patients to perform: " Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis. " Cardiac imagin

DateJan 30, 2016 (initiated Nov 3, 2015)
StatusTerminated
ReasonThe use of Allura Xper, Allura Clarity, Allura CV, Allura Centron and UNIQ systems regarding live images and still images can lead to still images being interpreted as live Images.
Root causeDevice Design
ActionThe firm, Philips, issued "URGENT-Field Safety Notices" dated 10/26/2015 and 11/19/2015 to its customers. The notices described the product, problem and actions to be taken. The customers were instructed to ensure the audible signal (X-ray buzzer) is always enabled, to activate only ONE fluoroscopy pedal at time, and to configure the footswitch to the default Philips configuration. Customers were also instructed to review the information with all members of your staff who need to be aware; retain a copy with the equipment IFU and complete and return the Confirmation Respond form to Mail Stop 4
Quantity13297 in total
DistributionWorldwide Distribution: US (Nationwide) including Puerto Rico and countries of: Albania, Algeria, US Virgin Islands, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Demo
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141979ALLURA XPER FD SERIES / ALLURA XPER OR TABLE SERIES
Philips Medical Systems Netherland BV, Aug 19, 2014