ClearanceBrief
Recall Z-0377-2016 ยท KDD Kit, Surgical Instrument, Disposable

Windstone Medical Packaging, Inc.

AMS 4675 Angio Pack. For use in a general clinical procedure.

DateNov 2, 2015
StatusTerminated
ReasonAMS 4675 Angio Pack is recalled because this kit containing U-SDN 2 Part Ultra Modified Seldinger Needle which is recalled by supplier.
Root causeProcess control
ActionAligned Medical Solutions (AMS) sent the Procedure Products -Urgent Medical Device Recall Letter, dated November 02, 2015 to consignees. This letter listed the items affected by recall as: AMS 4674 Custom Pack and AMS 4675 Angio Pack. Both direct account and sub account are informed that the part # NS/37 -0387 01 U-SDN 2 Part Ultra Modified Seldinger Needle (introducer needle) is recalled because some was unable to accept guide wires up to .038" as intended. The inner diameter was found to be out of specification in some cases. All other components in these two AMS kits are not affected b
Quantity40 kits
DistributionDistributed in AL and VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.