Recall Z-0343-2016 ยท DZC Wire, Orthodontic

Ormco/Sybronendo

Copper NiTi Preformed Archwires; Copper NiTi 40 BAF .016X.022 LO Pk10, Part No. 211-0952. Intended to provide tooth movement.

DateNov 30, 2015 (initiated Nov 4, 2015)
StatusTerminated
ReasonOrmco Corporation is recalling Copper NiTi Preformed Archwires because the product transformation temperature as listed on the labeling is incorrect.
Root causeEmployee error
ActionA recall letter dated 11/5/15 was sent to all their customers who purchased the Copper NiTi Preformed Archwires because it has been brought to Ormco's attention that the product transformation temperature as listed on the labeling is incorrect. The wire is labeled as having 40 degrees Celsius transformation temperature, but the actual wire has a transformation temperature of 29 degrees Celsius. The letter provides the customers with the problems identified and the actions to be taken. Customers are instructed to complete and fax back the enclosed acknowledgement form.
Quantity161 units
DistributionDistributed US (nationwide) and the countries of Canada, Japan, Mexico, Russia, and Peru.