Recall Z-0614-2016 ยท NQD Cardiac C-Reactive Protein, Antigen, Antiserum, And Control

Siemens Healthcare Diagnostics Inc

Siemens Acute Care" CCRP TestPak-in vitro diagnostic product intended to be used for calibration of the CCRP method on the Stratus CS analyzer Catalog Number: CCCRP SMN:10445066

DateJan 7, 2016 (initiated Nov 11, 2015)
StatusTerminated
ReasonStratus CS Cartridge Barcode labels may not scan on the Stratus CS systems
Root causeLabeling design
ActionSiemens Healthcare Diagnostics issued an Urgent Field Safety Notice by e-mail to all affected Siemens Healthcare Diagnostics Regional offices on November 11, 2015 both in the United States and internationally for communication with affected customers. This notice informs customers of the issue and provides mitigation instructions. Customers with questions were advised to contact their Siemens Customer Care Center or their local Siemens technical support representative. For questions regarding call 781-269-3000.
Quantity78
DistributionWorldwide Distribution - US (nationwide), Algeria, Armenia, Australia, Austria, Bangladesh, Bosnia Herzegovina, Canada, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Ireland, Italy, Japan, Libya, Malaysia, Poland, Portugal, Romania, Saudi Arabia, Slovenia, South Africa,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062924STRATUS CS ACUTE CARE CARDIOPHASE HSCRP TESTPAK
Dade Behring, Inc., Dec 26, 2006