Recall Z-0764-2016 ยท OWB Interventional Fluoroscopic X-Ray System

Philips Electronics North America Corporation

Philips Allura Xper FD20 Ceiling version only; 722028; imaging applications.

DateFeb 5, 2016 (initiated Nov 3, 2015)
StatusTerminated
ReasonNormally, movement of the C-arc is initially stopped by the motor (holding torque) then by the electromagnetic brake, leading to a full stop of the C-arc. Because of improper adjustment and tolerances, when the C-arc is initially stopped in an unbalanced position, the air gap of the brake can become too large, resulting in continued C-arc movement (C-arc is looking for balanced position).
Root causeDevice Design
ActionThe letters for the customers in the United States were sent on 11/3/2015, via certified return receipt and were addressed to the Risk Manager. Philips will perform a test on the 15 affected Allura Xper R8.2 FD 20C systems to see if the motion controller amplifier of the brake motor works correctly. If the test will not pass, the motor will be replaced. The Recall Package is sent to all Key Markets worldwide. The Key Market of an affected country will contact the facility where these systems were installed.
Quantity15 subject to correction and removal
DistributionWorldwide Distribution. US states of TN and IN and the countries of Australia, Belgium, Egypt, France, Germany, Japan, Republic of Korea, Netherlands, and Saudi Arabia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033737ALLURA XPER FD20
Philips Medical Systems North America, Inc., Dec 9, 2003